Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Heliyon ; 9(8): e19065, 2023 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-37636476

RESUMO

Purpose: Few studies have evaluated real-world performance of radiological AI-tools in clinical practice. Over one-year, we prospectively evaluated the use of AI software to support the detection of intracranial large vessel occlusions (LVO) on CT angiography (CTA). Method: Quantitative measures (user log-in attempts, AI standalone performance) and qualitative data (user surveys) were reviewed by a key-user group at three timepoints. A total of 491 CTA studies of 460 patients were included for analysis. Results: The overall accuracy of the AI-tool for LVO detection and localization was 87.6%, sensitivity 69.1% and specificity 91.2%. Out of 81 LVOs, 31 of 34 (91%) M1 occlusions were detected correctly, 19 of 38 (50%) M2 occlusions, and 6 of 9 (67%) ICA occlusions. The product was considered user-friendly. The diagnostic confidence of the users for LVO detection remained the same over the year. The last measured net promotor score was -56%. The use of the AI-tool fluctuated over the year with a declining trend. Conclusions: Our pragmatic approach of evaluating the AI-tool used in clinical practice, helped us to monitor the usage, to estimate the perceived added value by the users of the AI-tool, and to make an informed decision about the continuation of the use of the AI-tool.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...